Procedure & Laboratory Manual (continued from previous page)
LABORATORY 7. Place the verified wax try-in on the cast and make a silicone or plaster matrix of the tooth positions. Do not remove the teeth inside the matrix. If requesting fabrication of a virtual design CAM StructSURE® Bar or Framework, go to step 11.
If requesting fabrication of a Copy Milled CAM StructSURE Bar or Framework, go to the next step.
COPY MILLED CAM StructSURE 8. Place the verification index on the implant or abutment analogs and screw into place with waxing screws. Reduce the height of the temporary cylinders using a carbide bur so these fit within the confines of the silicone matrix. Apply a separator to the inside of the matrix of the wax try in and place it on the cast. Pour wax into the impressions of the teeth in the matrix and around the verification index. Remove the matrix and complete the wax up of the bar or framework on the lingual or palatal side.
9. Flask the wax up for processing. Boil away the wax and separate the flask. Remove the verification index and remove the temporary cylinders from the index. Place each Temporary Cylinder back on the analogs with waxing screws. Process the wax up in acrylic resin.
RESTORATIVE DENTIST - 2nd Visit 10. Remove the framework from the flask. Finish the framework to the desired contours with a bur. Cut back the teeth approximately 2mm if the framework is being designed for porcelain application. Place the matrix on the cast periodically as a guide for the proper contouring.
11. Complete the CAM StructSURE® Precision Milled Bar Work Order Form. See pages 7–8 for an example. Package the following items securely in a box:
Copy Of The Work Order Form
Verified Soft-tissue Master Cast
Verified Wax Try-In or Acrylic Resin framework
Verification Index
PLEASE DO NOT SEND:
The Articulator
The Opposing Cast
NOTE: All items and/or materials that have been used intraorally must be disinfected following manufacturer’s instructions before these are sent to BIOMET 3i.
Send to: In USA: BIOMET 3i ARCHITECH PSR® 4555 Riverside Drive Palm Beach Gardens Florida 33410
In Canada: BIOMET 3i Canada, Inc. 5805 St. Francois St. Laurent, Québec CANADA H4S 1B6 1-800-363-1980 Ext. 230
In Europe: BIOMET Spain Orthopaedics, S.L. ARCHITECH PSR® Calle Islas Baleares, 50 46988 Fuente del Jarro Valencia, Spain 34-96-137-95-00